Cefefe Stefenny Jolin1, Nurul Izzati Hamzan1*, Syarifah Nurhikmah Izzati Syed Nasarudin2, Siti Nurwani Ahmad Ridzuan1, Muhammad Nursyazwan Zamre1, Noor Hafizah Hassan1, Anasufiza Habib2
1. Special Protein Unit, Specialised Diagnostic Centre, Institute for Medical Research, Jalan Pahang, 50588 Kuala Lumpur, National Institutes of Health, Ministry of Health Malaysia, Malaysia
2. Specialised Diagnostic Centre, Institute for Medical Research, Jalan Pahang, 50588 Kuala Lumpur, National Institutes of Health, Ministry of Health Malaysia, Malaysia
3. Alfa Diagnostik Sdn Bhd (Co. No. 1033482-X) No.13, Jalan Delta U6/18, Taman Perindustrian, Sunway Subang, Seksyen U6, 40150 Shah Alam, Selangor, Malaysia
*Corresponding author: Nurul Izzati Hamzan.
ABSTRACT
INTRODUCTION: Venous blood collection tubes are critical for safe and accurate diagnostic sample collection. However, mechanical failures such as breakage during handling or accidental drops can pose risks, including biohazard exposure, sample loss, and diagnostic delays. Despite adherence to manufacturing standards, real-world evaluation of tube durability remains essential for clinical safety and quality assurance.
OBJECTIVE(S): This study aimed to evaluate the breakage resistance of IMPROVACUTER® venous blood collection tubes under standardised drop conditions, in compliance with ISO 6710:2017 guidelines.
MATERIALS & METHODS: A total of 72 IMPROVACUTER® tubes including Gel & Clot Activator, Clot Activator, EDTA.K2, three types of Citrate (3.2% single wall, 3.2% double wall, 3.8%), Heparin, and Glucose (NaF + EDTA) were filled to specified volumes and subjected to drop testing from a height of 1 metre onto a non-resilient (concrete) surface. Each tube underwent three replicate drops. Tubes were visually inspected for cracks, leaks, or breakage. A minimum 90% pass rate was required to meet ISO acceptance criteria.
RESULTS: All 72 tubes underwent thorough visual inspection for cracks, leaks, or structural breakage. Each tube successfully endured the drop tests without any damage, attaining a 100% acceptance rate.
CONCLUSION: IMPROVACUTER® venous blood collection tubes demonstrated excellent mechanical durability and full compliance with ISO standards under sudden mechanical shock conditions representative of worst-case handling scenarios. These results support the product’s reliability, enhance user confidence, and contribute to safer laboratory practices by minimising risks associated with tube breakage.
KEYWORDS: IMPROVACUTER®, Collection Tubes, Breakage, Accidental Drops, Laboratory Practices