Mohd Azerulazree Jamilan1, Nur Adriana Athirah Muhammad Rhadi1, Affandi Omar2, Aswir Abd Rashed1, Mohd Fairulnizal Md Noh3, Wan Ahmad Syazani Mohamed1*
1. Nutrition Unit, Nutrition, Metabolism & Cardiovascular Research Centre, Institute for Medical Research, National Institute of Health, Ministry of Health Malaysia, No. 1, Jalan Setia Murni U13/52, Setia Alam, 40170 Shah Alam, Selangor, Malaysia
2. Inborn Errors of Metabolism & Genetic Unit, Nutrition, Metabolism & Cardiovascular Research Centre, Institute for Medical Research, National Institute of Health, Ministry of Health Malaysia, No. 1, Jalan Setia Murni U13/52, Setia Alam, 40170 Shah Alam, Selangor, Malaysia
3. Nutrition, Metabolism & Cardiovascular Research Centre, Institute for Medical Research, National Institute of Health, Ministry of Health Malaysia, No. 1, Jalan Setia Murni U13/52, Setia Alam, 40170 Shah Alam, Selangor, Malaysia
*Corresponding author: Wan Ahmad Syazani Mohamed.
ABSTRACT
INTRODUCTION: Vitamin D plays a crucial role in human metabolism, particularly in promoting calcium absorption for bone mineralization and preventing hypocalcemic tetany. However, the body requires vitamin D to remain within an optimal range, as its deficiency can lead to bone-related diseases while excess levels may result in toxicity. Therefore, a reliable and cost-effective diagnostic tool is essential for monitoring and maintaining appropriate vitamin D levels.
OBJECTIVE(S): This study aimed to enhance the currently available commercial kit by making it more cost-effective, sensitive, and capable of establishing a detection range for more accurate quantification.
MATERIALS & METHODS: An improved method, based on a commercially available kit (25-OH Vitamin D3 HPLC Assay, Eagle Bioscience, Nashua, NH, USA), was used to extract plasma samples prior to Ultra-High-Performance Liquid Chromatography (UHPLC, Thermo Scientific, Waltham, MA, USA) analysis. The UHPLC was coupled with a diode array detector to observe the separation of 25OHDs peaks for quantification and method improvements. Several changes were made to improve the overall performance of the method. To reduce the cost, 70% (v/v) acetonitrile in deionized water was used as the mobile phase, replacing the provided mobile phase. Two of the provided columns (125 mm × 4 mm) were attached in line to the UHPLC with sample injection was changed from 50 μL to 100 μL in a 30-minute run, which was monitored at 264 nm.
RESULTS: A more cost-effective isocratic mobile phase using 70% (v/v) acetonitrile in deionized water enabled to retain of internal standard (8.6 min), 25OHD2 (21.9 min) and 25OHD3 (17.7 min) peaks in an excellent peak separation. It was found that the peak area of the 25OHD2 increased 4.2-fold, whereas 25OHD3 improved 4.5-fold when tested using the provided calibrator (CAL) and control (CTRL) samples. A linear correlation was observed for 25OHD2 (R2=0.9895; y=0.002x-0.0226) and 25OHD3 (R2=0.9856; y=0018x-0.0523) at 50 nM, 150 nM and 250 nM concentrations, respectively.
CONCLUSION: Improvement of the method enabled the protocol to be used in a diagnostic laboratory setting, yielding more cost-effective, sensitive, and reliable results, although a proper method validation was necessary to achieve this goal.
KEYWORDS: Vitamin D Metabolite, 25-Hydroxyvitamin D, Cost-effective Diagnostic Tool